Z-0837-2021 Class II Ongoing

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

Recall Details

Product Type
Devices
Report Date
January 20, 2021
Initiation Date
November 23, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5079 cases

Product Description

Naturalyte Liquid Acid concentrate bottles, Product Code 08-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason for Recall

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Distribution Pattern

US Nationwide distribution.

Code Information

Lot Numbers: 20LXAC058 20LXAC078 20LXAC096 20NXAC011 20NXAC012 Additional Affected lots as of 3/17/21: 20LXAC060 20NXAC026 20LXAC045 20LXAC043 20LXAC078 20NXAC012 20LXAC058 20LXAC096 20NXAC011 20LXAC059 20LXAC044 20LXAC079