Z-0837-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 20, 2021
- Initiation Date
- November 23, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 5079 cases
Product Description
Naturalyte Liquid Acid concentrate bottles, Product Code 08-3251-9 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Reason for Recall
The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.
Distribution Pattern
US Nationwide distribution.
Code Information
Lot Numbers: 20LXAC058 20LXAC078 20LXAC096 20NXAC011 20NXAC012 Additional Affected lots as of 3/17/21: 20LXAC060 20NXAC026 20LXAC045 20LXAC043 20LXAC078 20NXAC012 20LXAC058 20LXAC096 20NXAC011 20LXAC059 20LXAC044 20LXAC079