Z-0838-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- February 27, 2019
- Initiation Date
- September 25, 2017
- Termination Date
- January 21, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 142 units
Product Description
PROVEO 8 Ophthalmic Microscope System optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
Reason for Recall
The observed issues can lead to unexpected interruptions of the optics carrier focus drive movement, intermittent stops of the X-Y coupling movement or sporadic failures in the proper on-screen display functionality of the HDR recording unit. Additionally, an anomaly in the device software may also lead to the unexpected behavior that when a user releases a handle or foot switch button this may not immediately end the actuated functionality. This behavior can only occur in unlikely workflows where a user changes a user profile or a mode change is initiated at the same time as a functional button is activated.
Distribution Pattern
Worldwide
Code Information
Serial Numbers: 190216001; 060516001; 280916001; 061216001; 091216001; 301015001; 270516001; 180716001; 180716002; 190716001; 141016001; 211016001; 241016001; 261016001; 030317001; 060317001; 011216001; 020317001; 130716002; 100316001; 230216001; 041115001; 021115001; 270416001; 130516001; 160516001; 240516001; 220217001; 100417001; 131115001; 071216001; 100816001; 300816001; 231116001; 121216001; 100816001; 130616001; 081216001; 131216001; 140317001; 310816001; 160816001; 130117001; 060217001; 230317001; 280317001; 050417001; 170117001; 190416001; 070415004; 050516001; 090516001; 170516001; 020816001; 040816001; 111016001; 131116001; 150217001; 211116001; 090217001; 081116001; 020616001; 050117001; 031216001; 020416001; 010317001; 280416001; 271216001; 251115001; 250316001; 231216001; 230916001; 230217001; 260416001; 030616001; 080616001; 080316001; 250116001; 181016001; 220316001; 130716002; 201116001; 200217001; 180416001; 201216001; 170117002; 150116001; 121115001; 110316001; 060616001; 270217001; 090317001; 160317001; 190217001; 140716002; 101016001; 161115001; 190116001; 130116001; 010316001; 020217001; 121016001; 141016001; 161216001; 171216001; 180116001; 260516001; 300516001; 200716001; 260716001; 191116001; 171116001; 251116001; 131216001; 220416001; 281016001; 011116001; 270116001; 060416001; 231115001; 301015003; 301015002; 280116001; 150216001; 180216001; 290416001; 040516001; 190516001; 230516001; 250516001; 280616001; 300616001; 200616001; 230616001; 140716001; 130716001; 010916001; 050916001; 100117001; 260116001; 101116001; 041216001