Z-0839-2019 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North)

FDA device recall Z-0839-2019 was initiated by Randox Laboratories Ltd. on Dec 4, 2018 and is designated Class II. Reason for recall: Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results… The recall status is terminated (terminated Jun 18, 2020). Affected quantity: 225 kits.

Recall Details

Product Type
Devices
Report Date
Feb 27, 2019
Initiation Date
Dec 4, 2018
Termination Date
Jun 18, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
225 kits

Product Description

Randox CALCIUM (Ca) Colorimetric Method RX Series Cat. No. CA 3871, R1. Arsenazo Reagent 9 x 51ml GTIN: 05055273200904

Reason for Recall

Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.