Z-0841-2013 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-0841-2013 was initiated by Elekta, Inc. on Feb 11, 2013 and is designated Class II. Reason for recall: The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle. The recall status is terminated (terminated Jan 21, 2014). Affected quantity: 13.

Recall details

Product Type
Devices
Report Date
Feb 27, 2013
Initiation Date
Feb 11, 2013
Termination Date
Jan 21, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13

Product description

Monaco RTP System. Radiation treatment planning.

Reason for recall

The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.

Distribution pattern

Nationwide Distribution-including the states of AK, CA, FL, NJ, NC, OH, OR, TN, TX, WA, and WI.

Code information