Z-0841-2013 Class II Terminated
FDA device recall Z-0841-2013 was initiated by Elekta, Inc. on Feb 11, 2013 and is designated Class II. Reason for recall: The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle. The recall status is terminated (terminated Jan 21, 2014). Affected quantity: 13.
Recall details
- Product Type
- Devices
- Report Date
- Feb 27, 2013
- Initiation Date
- Feb 11, 2013
- Termination Date
- Jan 21, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13
Product description
Monaco RTP System. Radiation treatment planning.
Reason for recall
The beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.
Distribution pattern
Nationwide Distribution-including the states of AK, CA, FL, NJ, NC, OH, OR, TN, TX, WA, and WI.
Code information
Versions 3.20.00 and 3.20.01