Z-0855-2014 Class II Terminated

Recalled by Orthofix, Inc. — Lewisville, TX

FDA device recall Z-0855-2014 was initiated by Orthofix, Inc. on Nov 26, 2013 and is designated Class II. Reason for recall: Orthofix received 6 complaints which resulted in a reportable events due to extended surgical times greater than 30 minutes for the Modular Screw Driver (PN 52-1332). The complaints alleged that the… The recall status is terminated (terminated Dec 9, 2014). Affected quantity: 443.

Recall details

Product Type
Devices
Report Date
Feb 5, 2014
Initiation Date
Nov 26, 2013
Termination Date
Dec 9, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
443

Product description

Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone screw into the pedicle during a spinal fixation procedure. It is a reusable instrument.

Reason for recall

Orthofix received 6 complaints which resulted in a reportable events due to extended surgical times greater than 30 minutes for the Modular Screw Driver (PN 52-1332). The complaints alleged that the screw driver's collet would malfunction resulting in the surgeon being unable to use the Modular Screw Driver to effectively place Modular Screws, which may result in a delay of surgery.

Distribution pattern

Worldwide Distribution - US Nationwide including (Puerto Rico) and the countries of: Spain, Germany, Australia, Peru, Mexico, Colombia, South Africa, and Israel.

Code information

Production Identification Numbers: Part Number 52-1332, All Lot Numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0855-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Orthofix, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0855-2014. Report a data issue: hello@regdatalab.com.