Z-0856-2021 Class II Ongoing

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
January 20, 2021
Initiation Date
December 10, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
81 units

Product Description

Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

Reason for Recall

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Distribution Pattern

Worldwide

Code Information

Discovery NM 630 (Model Number H3101RH), Serial numbers: 63GY50110, 63GY50119, 63GY50116, 63GZ50141, 63GY50120, 63GY50108, 63GY50121, 63GZ50142, 63GY50112, 63GZ50139, 63GY50107, 63GZ50138 NM 830 (Model Number H3910AC), Serial Numbers: 830Z60199, 830Z60182, 830Z60183, 830Z60192, 830Y60089, 830Y60105, 830Y60079, 830Z60207, 830Y60092, 830Y60102, 830Y60081, 830Z60200, 830Y60106, 830Z60201, 83EZ88001, 830Y60111, 830Y60094, 830Z60211, 830Y60101, 830Z60210, 830Y60118, 830Z60195, 830Z60194, 830Z60198, 830Z60181, 830Z60202, 830Y60097, 830Z60196, 830Z60180, 830Y60078, 830Y60115, 830Y60114, 830Y60108, 830Y60100, 830Y60087, 830Y60084, 830Z60188, 830Z60206, 830Y60112, 830Y60082, 830Y60093, 830Z60191, 830Y60107, 830Y60080, 830Z60190, 830Y60103, 830Z60193, 830Y60095, 830Y60119, 830Z60209, 830Z60205, 830Y60073, 830Y60120, 830Y60083, 830Y60116, 830Z60203, 830Y60085, 830Y60077, 830Z60184, 830Y60086, 830Z60187, 830Y60113, 830Z60208, 830Y60110, 830Z60204, 830Y60072, 830Z60179, 830Z60186 830 NM Cyber Defense (Model Number H3910AG), Serial Numbers: 830Z60185