Z-0856-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 20, 2021
- Initiation Date
- December 10, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 81 units
Product Description
Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.
Reason for Recall
GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.
Distribution Pattern
Worldwide
Code Information
Discovery NM 630 (Model Number H3101RH), Serial numbers: 63GY50110, 63GY50119, 63GY50116, 63GZ50141, 63GY50120, 63GY50108, 63GY50121, 63GZ50142, 63GY50112, 63GZ50139, 63GY50107, 63GZ50138 NM 830 (Model Number H3910AC), Serial Numbers: 830Z60199, 830Z60182, 830Z60183, 830Z60192, 830Y60089, 830Y60105, 830Y60079, 830Z60207, 830Y60092, 830Y60102, 830Y60081, 830Z60200, 830Y60106, 830Z60201, 83EZ88001, 830Y60111, 830Y60094, 830Z60211, 830Y60101, 830Z60210, 830Y60118, 830Z60195, 830Z60194, 830Z60198, 830Z60181, 830Z60202, 830Y60097, 830Z60196, 830Z60180, 830Y60078, 830Y60115, 830Y60114, 830Y60108, 830Y60100, 830Y60087, 830Y60084, 830Z60188, 830Z60206, 830Y60112, 830Y60082, 830Y60093, 830Z60191, 830Y60107, 830Y60080, 830Z60190, 830Y60103, 830Z60193, 830Y60095, 830Y60119, 830Z60209, 830Z60205, 830Y60073, 830Y60120, 830Y60083, 830Y60116, 830Z60203, 830Y60085, 830Y60077, 830Z60184, 830Y60086, 830Z60187, 830Y60113, 830Z60208, 830Y60110, 830Z60204, 830Y60072, 830Z60179, 830Z60186 830 NM Cyber Defense (Model Number H3910AG), Serial Numbers: 830Z60185