Z-0857-2021 Class II Ongoing

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
January 20, 2021
Initiation Date
December 10, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
52 untits

Product Description

Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.

Reason for Recall

GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Distribution Pattern

Worldwide

Code Information

Discovery 670 DR (Model H3100BT), Serial Numbers: DRGY56033, DRGY56037, DRGY56035, DRGY56030, DRGZ56043, DRGZ56042, DRGY56032, DRGY56031, DRGY56034, DRGY56036 NM/CT 870 DR (Model 3100Q), Serial Numbers: 870Y64066, and (Model H3100AS), Serial Numbers: 870Y64077, 870Y64064, 870Y64081, 870Y64079, 870Y64067, 870Y64074, 870Z64117, 870Y64078, 870Y64080, 870Y64069, 870Y64085, 870Z64121, 870Z64119, 870Z64122, 870Y64056, 870Y64063, 870Z64114, 870Z64115, 870Y64068, 870Y64082, 870Z64116, 870Z64120, 870Y64065, 870Y64084, 870Y64061, 870Y64083, 870Z64113, 870Y64070, 870Z64123, 870Y64058, 870Z64118, 870Y64057 Discovery NM670 Pro (Model 3100XB), Serial Numbers: PRGY52027, PRGY52031, PRGZ52047, PRGZ52046, PRGY52028, PRGZ52048, PRGY52029, PRGY52030 Discovery 670 ES (Model H3100XF), Serial Numbers: ESGZ53005