Z-0864-2017 Class III Terminated

Recalled by Ortho-Clinical Diagnostics — Rochester, NY

FDA device recall Z-0864-2017 was initiated by Ortho-Clinical Diagnostics on Oct 26, 2016 and is designated Class III. Reason for recall: CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the… The recall status is terminated (terminated Jun 22, 2018). Affected quantity: 124,063 units.

Recall details

Product Type
Devices
Report Date
Dec 28, 2016
Initiation Date
Oct 26, 2016
Termination Date
Jun 22, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
124,063 units

Product description

VITROS CRP Slides: 1) VITROS¿ Chemistry Products CRP Slides (250 slides per sales unit), Catalog # 1926740 and 2) , VITROS¿ Chemistry Products CRP Slides (90 slides per sales unit), Catalog # 1926740

Reason for recall

CRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is not subtracted from the final result. The IFUs do not provide specific instructions for how to account for endogenous CRP in the diluent when calculating the sample result. Orthos diluents used to dilute samples containing CRP are manufactured using human plasma based fluids that contain endogenous CRP. Without accounting for the CRP in the diluent and removing its contribution from the result, the reported result will include the CRP present in both the sample and diluent, possibly leading to positively biased results.

Distribution pattern

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico, and to the countries of : Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain and Poland.

Code information

Affected Generations (GENs): 3, 11, 12, 30 through 36

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0864-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0864-2017. Report a data issue: hello@regdatalab.com.