Z-0866-2020 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm, N/A

Recall Details

Product Type
Devices
Report Date
February 5, 2020
Initiation Date
December 19, 2019
Termination Date
November 24, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3114 Devices/Licenses

Product Description

RayStation stand-alone software treatment planning system. Models 4.0, 4.3, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0.

Reason for Recall

Three issues found: i) The Map ROI options in the ROI list in the Structure Definition module may generate unintended ROI geometries ii) Elekta guard leaf behavior. There is an interoperability issue with Elekta regarding setting of guard leaves. iii)SSD, when intended as source-to-surface distance, it sometimes gives source-to-skin distance.To the best of our knowledge, these issues have not caused any patient mistreatment or other incidents. However, the user must be aware of the following information to avoid incorrect dose calculations during treatment planning.

Distribution Pattern

US Nationwide distribution in states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NV, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, WI

Code Information

Software Versions Numbers: 4.0.0.14 , 4.0.1.4 , 4.0.2.9 , 4.0.3.4 , 4.3.0.14 , 4.5.0.19 , 4.5.1.14 , 4.5.2.7 , 4.7.0.15 , 4.7.1.10 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.9.0.42 , 5.0.0.37 , 5.0.1.11 5.0.2.35 , 6.0.0.24 , 6.1.0.26 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 .