Z-0872-2024 Class II Ongoing

Recalled by Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-0872-2024 was initiated by Howmedica Osteonics Corp. on Dec 1, 2023 and is designated Class II. Reason for recall: Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties. The recall status is ongoing. Affected quantity: 70 units US.

Recall details

Product Type
Devices
Report Date
Feb 7, 2024
Initiation Date
Dec 1, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
70 units US

Product description

SCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032

Reason for recall

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Distribution pattern

Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.

Code information

GTIN: 07613327033823 Lot Numbers: 6LR4, R14V, N882

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0872-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0872-2024. Report a data issue: hello@regdatalab.com.