Z-0880-2020 Class II Terminated

Recalled by Synthes, Inc. — West Chester, PA

Recall Details

Product Type
Devices
Report Date
February 5, 2020
Initiation Date
December 16, 2019
Termination Date
June 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25 total

Product Description

2.4MM TI VA LOCKING SCREW STARDRIVE 12MM STERILE, Part Number 04.210.112TS - Product Usage:Sterile Tube Packaging is a packaging system to deliver single, sterile screws to the sterile field.

Reason for Recall

There is a possibility that the inner cap could open when the outer cap is removed or the entire inner tube remains in the outer tube. In both instances this could prevent the screw from being removed and used causing a surgical delay. The inner tube should remain closed and retained in the outer cap when presented to the sterile field.

Distribution Pattern

US Nationwide distribution in the states of AL and ID.

Code Information

Lot Number: 6L36824