Z-1799-2015 Class II Terminated

Recalled by Synthes, Inc. — West Chester, PA

FDA device recall Z-1799-2015 was initiated by Synthes, Inc. on May 28, 2015 and is designated Class II. Reason for recall: Certain affected parts and lots of the Pull Reduction Devices may have been manufactured to an incorrect hardness specification which could result in intraoperative breakage. The recall status is terminated (terminated Aug 22, 2016). Affected quantity: 3747 US, 19 Foreign.

Recall details

Product Type
Devices
Report Date
Jun 24, 2015
Initiation Date
May 28, 2015
Termination Date
Aug 22, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3747 US, 19 Foreign

Product description

Pull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling is placed through the guide sleeve and plate holes to pull or push bone fragments relative to the plate. This instrument can be used for: Minor Varus-valgus adjustments (approximately 2¿ - 4¿), Translational adjustments, Stabilization of plate bone orientation during insertion of the first screws, alignment of segmental fragments and pre-drilling dense or thick cortical bone before placing a 5.0 mm locking screw.

Reason for recall

Certain affected parts and lots of the Pull Reduction Devices may have been manufactured to an incorrect hardness specification which could result in intraoperative breakage.

Distribution pattern

Worldwide Distribution-US (nationwide) and internationlly to Canada

Code information

Part Number: 03.120.023, with lot numbers: 1825699 1825946 1959344 1959346 3052800 3066995 3066996 3066997 3092693 3092737 3139434 3139480 3139483 3139485 3159152 3167363 3172147 3207364 3218030 3218068 3223545 3224270 3224271 3224275 3224581 3224582 3224583 3233445 3288206 3422091 3447606 3469240 3502469 3535235 3556421 3556422 3584316 3642892 3740917 3740919 3740920 3740924 3775480 3775482 3811200 7510644 7510645 7510647 7510648 7542178 7555756 7629268 7699116 7699117 7742370 7742371 7742372 7778806 7778807 7821741 7878176 7995396 8016667 8028575 8068778 8070887 8070891 8195690 8210669 8217914 8408715

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1799-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Synthes, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1799-2015. Report a data issue: hello@regdatalab.com.