Z-0884-2013 Class II Terminated

Recalled by Philips Healthcare Inc. — Andover, MA

FDA device recall Z-0884-2013 was initiated by Philips Healthcare Inc. on Oct 25, 2012 and is designated Class II. Reason for recall: When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the… The recall status is terminated (terminated Mar 31, 2014). Affected quantity: 551.

Recall details

Product Type
Devices
Report Date
Mar 6, 2013
Initiation Date
Oct 25, 2012
Termination Date
Mar 31, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
551

Product description

Philips DigitalDiagnost X-Ray System with Eleva Software 2.1.3 Model numbers 712020, 712022, and 712082 Product Usage: Digital radiographic system

Reason for recall

When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.

Distribution pattern

Worldwide Distribution - USA Nationwide and the Foreign countries of: Austria, Canada, Belgium, China,Denmark, France, germany, Italy, Netherlands,New Zealand, Norway, Philippines, Portugual, SOuth Korea, Spain, Switzerland, Thailand, and UK.

Code information

Systems with Eleva SW version 2.1.3

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0884-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Healthcare Inc. →

Source: openFDA record for Z-0884-2013. Report a data issue: hello@regdatalab.com.