Z-0888-2024 Class II Ongoing

Recalled by Howmedica Osteonics Corp. — Mahwah, NJ

Recall Details

Product Type
Devices
Report Date
February 7, 2024
Initiation Date
December 1, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
269 units US; 307 units OUS

Product Description

SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-0710

Reason for Recall

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Distribution Pattern

Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.

Code Information

GTIN: 07613327050295 Lot Numbers: RLAR, 2P84, EW6V, 90AV, 4JHN, 65A3, 22MP, NNE7, Y373, 58LA, VXAL, 4201, A2AN, 5YP8, X0P9, 9A8T,6LN7, WEL3, NDK3, 25WJ, T2JW, YVWT, 7DR0, DMVL,