Z-0891-2020 Class II Terminated

Recalled by Haemonetics Corporation — Braintree, MA

Recall Details

Product Type
Devices
Report Date
February 5, 2020
Initiation Date
October 21, 2019
Termination Date
November 6, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
690043

Product Description

Cell Saver¿5/5+ Bowl Kit (225 ml) Autotransfusion Device - Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.

Reason for Recall

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.

Code Information

Catalog number/FG Item Number: 00263-00 1. Specific Lots with Bowl Cracks (44831 Distributed) Lot number: 0119010 0119011 0219031 0218056 0218057 0218058 0418012 1018084 1118009 0718099 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.