Z-0903-2020 Class II Terminated

Recalled by Haemonetics Corporation — Braintree, MA

FDA device recall Z-0903-2020 was initiated by Haemonetics Corporation on Oct 21, 2019 and is designated Class II. Reason for recall: Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a… The recall status is terminated (terminated Nov 6, 2020). Affected quantity: 21113.

Recall details

Product Type
Devices
Report Date
Feb 5, 2020
Initiation Date
Oct 21, 2019
Termination Date
Nov 6, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21113

Product description

JAPAN CELL SAVER ELITE SET - 125ML - Product Usage: its related accessory components are intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient.

Reason for recall

Potential to develop leaks at the inner core of Cell Saver¿ 5/5+ and Cell Saver¿ Elite¿/Elite¿ +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.

Code information

Catalog number/FG Item Number: CSE-P-125-JA 1. No Specific Lots with Bowl Cracks (0 Distributed) 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0903-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Haemonetics Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0903-2020. Report a data issue: hello@regdatalab.com.