Z-0915-2024 Class II Ongoing

Recalled by Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-0915-2024 was initiated by Howmedica Osteonics Corp. on Dec 1, 2023 and is designated Class II. Reason for recall: Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties. The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Feb 7, 2024
Initiation Date
Dec 1, 2023
Voluntary/Mandated
Voluntary: Firm initiated

Product description

LARGE PRIMARY REV.TIB.INSERT- DURACON Intended for knee replacement Product Number: 6637-4-231

Reason for recall

Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Distribution pattern

Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.

Code information

GTIN: 04546540307101 Lot Numbers: 33164901

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0915-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0915-2024. Report a data issue: hello@regdatalab.com.