Z-0918-2025 Class II Ongoing

Recalled by GE Medical Systems, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
January 22, 2025
Initiation Date
December 6, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 systems

Product Description

Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB

Reason for Recall

There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.

Code Information

UDI-DI: 00195278659446, Serial Numbers: SWT24040001TA SWT24090006TA SWT24150008TA SWT24050001TA SWT24180011TA SWT24180009TA SWT31240012TA SWT24340007TA SWT24360002TA SWT24370001TA SWT24370002TA SWT24180010TA 6399074-7 6399074-8 6399074-9 6399326-1 6399326-2 SWT24060001TA