Z-0918-2025 Class II Ongoing
FDA device recall Z-0918-2025 was initiated by GE Medical Systems, LLC on Dec 6, 2024 and is designated Class II. Reason for recall: There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time. The recall status is ongoing. Affected quantity: 7 systems.
Recall details
- Product Type
- Devices
- Report Date
- Jan 22, 2025
- Initiation Date
- Dec 6, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 systems
Product description
Reason for recall
Distribution pattern
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0918-2025
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026