Z-0926-2018 Class II Terminated

Recalled by Abbott Point Of Care Inc. — Princeton, NJ

Recall Details

Product Type
Devices
Report Date
March 14, 2018
Initiation Date
September 12, 2017
Termination Date
August 23, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
894

Product Description

i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)

Reason for Recall

Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.

Distribution Pattern

Distributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.

Code Information

All units that have been upgraded to software version 2.8