Z-1999-2026 Class II Ongoing

Recalled by Abbott Point Of Care Inc. — Princeton, NJ

Recall Details

Product Type
Devices
Report Date
May 6, 2026
Initiation Date
April 1, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
369650 units (150875 US, 218775 OUS)

Product Description

i-STAT EG7+ cartridge; List Number: 03P76-25;

Reason for Recall

Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.

Code Information

List Number: 03P76-25; UDI-DI: 00054749000043; Lot numbers: N25206, N25207, N25209, N25213, N25218, N25219, N25233;