Z-0960-2025 Class I Ongoing

Recalled by Percussionaire Corporation — Sandpoint, ID

FDA device recall Z-0960-2025 was initiated by Percussionaire Corporation on Dec 23, 2024 and is designated Class I. Reason for recall: Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation. The recall status is ongoing. Affected quantity: 4,727 packs/cases.

Recall details

Product Type
Devices
Report Date
Jan 29, 2025
Initiation Date
Dec 23, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,727 packs/cases

Product description

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients

Reason for recall

Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

Distribution pattern

Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.

Code information

-Phasitron Kit, VDR, Single Patient, 5pk Model Number - A50094-D-5PK UDI-DI code: 00849436000259 Lot Codes: WO04294 WO04424 WO04764 WO05070 WO05186 WO05460 WO05685 WO05910 WO06388 WO06576 WO06883 WO07095 WO07196 WO07317 WO07405 WO07450 WO07696 - Control Unit Tester Model Number: A51001 VDR4 Lot Codes: WO04750 WO06701 - VDR W/Swivel T Single Patient Phasitron Model Number: PRT 991 Lot Code: WO04733 - VDR4 Humidification Adapter Kit, Case of 10 Model Number: PRT 992 Lot Code: WO045667 - VDR4 Humidification Kit with Cross Tee, Case of 10 Model Number: PRT 993 Lot Codes: WO04745 WO04832 WO04893 WO04893 WO05309 WO06523 WO07080 WO07283

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0960-2025
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Percussionaire Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Percussionaire Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Percussionaire Corporation →

Source: openFDA record for Z-0960-2025. Report a data issue: hello@regdatalab.com.