Z-0960-2025 Class I Ongoing
FDA device recall Z-0960-2025 was initiated by Percussionaire Corporation on Dec 23, 2024 and is designated Class I. Reason for recall: Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation. The recall status is ongoing. Affected quantity: 4,727 packs/cases.
Recall details
- Product Type
- Devices
- Report Date
- Jan 29, 2025
- Initiation Date
- Dec 23, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,727 packs/cases
Product description
Reason for recall
Distribution pattern
Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0960-2025
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026