Z-3324-2026 Class II Ongoing

Recalled by Percussionaire Corporation — Sandpoint, ID

FDA device recall Z-3324-2026 was initiated by Percussionaire Corporation on Aug 18, 2026 and is designated Class II. Reason for recall: Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile… The recall status is ongoing. Affected quantity: 228,596.

Recall details

Product Type
Devices
Report Date
Oct 7, 2026
Initiation Date
Aug 18, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
228,596

Product description

Travel Air, Product Number: S00055; Phasitron 5, Product Number: P5 used with Instruction Manual

Reason for recall

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Distribution pattern

Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.

Code information

Product Number/UDI-DI: S00055/00849436000242, P5/00849436000501, Instruction Manual: P20106 Rev B

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3324-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-3324-2026. Report a data issue: hello@regdatalab.com.