Z-0965-2014 Class III Terminated

Recalled by Bio-Rad Laboratories, Inc. — Hercules, CA

FDA device recall Z-0965-2014 was initiated by Bio-Rad Laboratories, Inc. on Nov 22, 2013 and is designated Class III. Reason for recall: D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the… The recall status is terminated (terminated Mar 20, 2014). Affected quantity: 2906 kits (318 Kits in US; 2588 kits internationally).

Recall details

Product Type
Devices
Report Date
Feb 19, 2014
Initiation Date
Nov 22, 2013
Termination Date
Mar 20, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2906 kits (318 Kits in US; 2588 kits internationally)

Product description

D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.

Reason for recall

D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Latin America, Asia Pacific, Canada, Australia, New Zealand and India.

Code information

Catalog number: 220-0201: Reorder Pack Lot No./Expiration date: 20030322 3/31/2014 20030337 3/31/2014 64000081 5/31/2014 64000331 5/31/2014 64000332 5/31/2014 64000888 08/31/2014 64001344 04/30/2014 64002239 11/30/2014 ( 64002240 08/3112014 64002241 11/30/2014

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0965-2014
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0965-2014. Report a data issue: hello@regdatalab.com.