Z-0973-2014 Class II Terminated

Recalled by Medtronic Navigation, Inc. — Littleton, MA

FDA device recall Z-0973-2014 was initiated by Medtronic Navigation, Inc. on Jul 25, 2013 and is designated Class II. Reason for recall: Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for… The recall status is terminated (terminated Apr 17, 2019). Affected quantity: 449 in US.

Recall details

Product Type
Devices
Report Date
Mar 12, 2014
Initiation Date
Jul 25, 2013
Termination Date
Apr 17, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
449 in US

Product description

Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems.

Reason for recall

Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)

Distribution pattern

USA (nationwide) Distribution

Code information

Model No. BI-700-00027 and BI-700-00028 Serial number range: 00102 through 00753 and B13080721, B13080722, B13090729, B13120763, B13120764, B13120766, B13120768, B13120769

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0973-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Navigation, Inc. →

Source: openFDA record for Z-0973-2014. Report a data issue: hello@regdatalab.com.