Z-0976-2018 Class II Terminated

Recalled by Deerfield Imaging, Inc. — Minnetonka, MN

Recall Details

Product Type
Devices
Report Date
March 21, 2018
Initiation Date
February 20, 2018
Termination Date
August 27, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 tables

Product Description

IMRIS T2X Table, Part Number 109682-600

Reason for Recall

The rotational locking mechanisms on the ORT 100/ T2X tables can become disengaged or unlocked while in use which could potentially result in unintended movement of the table.

Distribution Pattern

US, Canada, Australia, China, Qatar

Code Information

Serial Numbers: 10000239, 10001306, 10000983, 10000982, 10000233, 10001119, 10001192, 10002375, 10003623, 10003623, 05 002, 10003634, 10001118, 10001188, 10001340, 10003638