Z-0993-2015 Class II Terminated
FDA device recall Z-0993-2015 was initiated by Maquet Medical Systems USA on Aug 21, 2014 and is designated Class II. Reason for recall: The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS… The recall status is terminated (terminated Nov 17, 2015). Affected quantity: 6 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 28, 2015
- Initiation Date
- Aug 21, 2014
- Termination Date
- Nov 17, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6 units
Product description
TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet Tegris OR Integration System is designed to be used as the central operating unit in an operating room. The integration system has two main functions: Recording and distribution of images and videos and interaction with medical and non-medical devices.
Reason for recall
The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.
Distribution pattern
US Nationwide Distribution including NE, NY and PA.
Code information
part number: 1703.35AO, with software version up to 2.1.x.x.