Z-0993-2015 Class II Terminated

Recalled by Maquet Medical Systems USA — Wayne, NJ

FDA device recall Z-0993-2015 was initiated by Maquet Medical Systems USA on Aug 21, 2014 and is designated Class II. Reason for recall: The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS… The recall status is terminated (terminated Nov 17, 2015). Affected quantity: 6 units.

Recall details

Product Type
Devices
Report Date
Jan 28, 2015
Initiation Date
Aug 21, 2014
Termination Date
Nov 17, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product description

TEGRIS System manufactured by MAQUET GMBH in Germany The Maquet Tegris OR Integration System is designed to be used as the central operating unit in an operating room. The integration system has two main functions: Recording and distribution of images and videos and interaction with medical and non-medical devices.

Reason for recall

The wrong movement results on the MAGNUS OR table. The button commands on the TEGRIS touchscreen for Lower Leg Up and Upper Leg Down are switched in the software for the integration with the MAGNUS operating table system.

Distribution pattern

US Nationwide Distribution including NE, NY and PA.

Code information

part number: 1703.35AO, with software version up to 2.1.x.x.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0993-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0993-2015. Report a data issue: hello@regdatalab.com.