Z-1003-2013 Class II Terminated

Recalled by Philips Healthcare Inc. — Andover, MA

FDA device recall Z-1003-2013 was initiated by Philips Healthcare Inc. on Apr 5, 2010 and is designated Class II. Reason for recall: Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated. The recall status is terminated (terminated Mar 8, 2021). Affected quantity: 2089 US 4036 ROW.

Recall details

Product Type
Devices
Report Date
Apr 24, 2013
Initiation Date
Apr 5, 2010
Termination Date
Mar 8, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2089 US 4036 ROW

Product description

Philips Healthcare Computed Tomography X-Ray System. These devices are whole-body computed tomography (CT) x-ray systems or sub-systems, each with a continuously rotating x-ray tube and multi-row detectors enclosed by a gantry. X-ray transmission data acquired and taken at different angles can be reconstructed into cross-sectional images. Each device also includes signal analysis and display equipment, patient and equipment supports, components, and accessories.

Reason for recall

Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.

Distribution pattern

Worldwide Distribution including USA (nationwide)

Code information

Gemini Dual, Gemini 16 Power, Gemini GXL, Gemini LXL, Gemini TF 16, Gemini TF Base/Ready, Gemini TF 64, Gemini TF Big Bore, Precedence SPECT/CT, Brilliance CT (6, 10,16, 16P, 40, 64, and Big Bore), Brilliance iCT, Brilliance iCT SP

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1003-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Healthcare Inc. →

Source: openFDA record for Z-1003-2013. Report a data issue: hello@regdatalab.com.