Z-1004-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 20, 2019
- Initiation Date
- June 21, 2018
- Termination Date
- September 4, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 45
Product Description
Ossur, LIL Angel Vest, Part/ Description: 545100K/ LIL ANGEL VEST P2 KODEL; 545100L/ LIL ANGEL VEST P2 LAMBSWOOL; 545100S/ LIL ANGEL VEST P2 SORBATEX; 545150K/ LIL ANGEL VEST P3 KODEL; 545150L/ LIL ANGEL VEST P3 LAMBSWOOL; 545150S/ LIL ANGEL VEST P3 SORBATEX
Reason for Recall
Halo system products labeled MR conditional will undergo confirmatory safety testing to confirm MR conditional parameters. In addition, the second line description on the titanium pin label mistakenly lists ceramic tip. The second line description should state titanium pin.
Distribution Pattern
U.S.: HI, AL, CA, WA, AZ, TX, NM, NE, MI, IL, MN, MS, IN, OH, KY, TN, AK, AL, UT, SC, NC, GA, PA, FL, NY, DE, RI, VA, MA; Foreign: Australia, Germany, New Zealand, Belgium
Code Information
All lots of the products manufactured from 06/30/2009 - 12/13/2018