Z-1014-2022 Class II Ongoing

Recalled by Instrumentation Laboratory — Bedford, MA

FDA device recall Z-1014-2022 was initiated by Instrumentation Laboratory on Feb 22, 2022 and is designated Class II. Reason for recall: Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa… The recall status is ongoing. Affected quantity: 2725 units US: 5,572 units OUS.UPDATE: 2727 units US; 5277 units OUS.

Recall Details

Product Type
Devices
Report Date
May 4, 2022
Initiation Date
Feb 22, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2725 units US: 5,572 units OUS.UPDATE: 2727 units US; 5277 units OUS

Product Description

ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS 00000280060R (Refurbished) 3. ACL TOP 350 CTS 00000280065 4. ACL TOP 500 CTS 00000280040 5. ACL TOP 500 CTS 00000280040R ( Refurbished) 6/ 6. ACL TOP 550 CTS 00000280045 7. ACL TOP 970 CL 00000280097

Reason for Recall

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, El Salvador, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, S. Tome, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad, Tobago, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam.

Code Information

All Serial Numbers. Trade Name Part Number UDI ACL TOP 300 CTS 00000280060 08426950556916; "ACL TOP 300 CTS (Refurbished)" 00000280060R 08426950928027; ACL TOP 350 CTS 00000280065 08426950784081; ACL TOP 500 CTS 00000280040 08426950453499; "ACL TOP 500 CTS (Refurbished)" 00000280040R 08426950928003; ACL TOP 550 CTS 00000280045 08426950729242; ACL TOP 970 CL 00000280097 08430793045476.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1014-2022
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.