Z-1017-2019 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

Recall Details

Product Type
Devices
Report Date
March 27, 2019
Initiation Date
December 3, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
141 kits

Product Description

Fructosamine Control 1, FR2994 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.

Reason for Recall

The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%.

Distribution Pattern

US Distribution to states to: CA, ME. MI, and WV.

Code Information

499FR