Z-1018-2019 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-1018-2019 was initiated by Randox Laboratories Ltd. on Dec 3, 2018 and is designated Class II. Reason for recall: The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and… The recall status is ongoing. Affected quantity: 133 kits.

Recall Details

Product Type
Devices
Report Date
Mar 27, 2019
Initiation Date
Dec 3, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
133 kits

Product Description

Fructosamine Control 3, FR2996 This product is intended for in vitro use in the calibration of the Randox Liquid Fructosamine assay on clinical chemistry systems.

Reason for Recall

The firm is conducting a Device Modification for Fructosamine Calibrator, Fructosamine Control 1 and Control 3 for the lots specified. The assigned value of Fructosamine in this calibrator and controls has now been re-assigned by a decrease of 25%.

Distribution Pattern

US Distribution to states to: CA, ME. MI, and WV.

Code Information

500FR

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1018-2019
openFDA file date
Oct 21, 2026
Loaded by RegDataLab
Oct 23, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.