Z-1057-2014 Class II Terminated
FDA device recall Z-1057-2014 was initiated by Synthes, Inc. on Jan 7, 2014 and is designated Class II. Reason for recall: The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate. The recall status is terminated (terminated Aug 18, 2016). Affected quantity: 143.
Recall details
- Product Type
- Devices
- Report Date
- Feb 26, 2014
- Initiation Date
- Jan 7, 2014
- Termination Date
- Aug 18, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 143
Product description
Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving Cap/Threaded is threaded into the insertion handle and secured with a wrench. If desired, the nail can then be inserted using light hammer blows.
Reason for recall
The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate.
Distribution pattern
USA (nationwide) Distribution.
Code information
Part No. 03.010.523, Lot Nos. 8279162 & 8429004
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1057-2014
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for Synthes, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.