Z-1082-2014 Class II Terminated
FDA device recall Z-1082-2014 was initiated by Elekta, Inc. on Nov 25, 2013 and is designated Class II. Reason for recall: During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value. The recall status is terminated (terminated Sep 15, 2021). Affected quantity: 1309.
Recall details
- Product Type
- Devices
- Report Date
- Mar 12, 2014
- Initiation Date
- Nov 25, 2013
- Termination Date
- Sep 15, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1309
Product description
Reason for recall
Distribution pattern
Worldwide Distribution: US (nationwide): AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IA, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, including Puerto Rico, DC; and countries of: Albania, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bosnia & Herzego, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Gabon, Germany, Ghana, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore Slovakia, South Africa, South Korea, Spain, Sudan, Suriname, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Ukraine, Venezuela, and Vietnam.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1082-2014
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026