Z-1084-2014 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-1084-2014 was initiated by Elekta, Inc. on Nov 25, 2013 and is designated Class II. Reason for recall: During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value. The recall status is terminated (terminated Nov 8, 2017). Affected quantity: 300.

Recall details

Product Type
Devices
Report Date
Mar 12, 2014
Initiation Date
Nov 25, 2013
Termination Date
Nov 8, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
300

Product description

Agility. Version 3.0 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators which is associated with Agility collimators.

Reason for recall

During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.

Distribution pattern

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AK, AZ, CA, CT, FL, GA, IL, KY, LA, MD, MA, MI, MO, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, and WY, and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Czech Republic, Cuba, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan , Panama, Peru, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Ukraine, Venezuela, and Vietnam.

Code information

Model 3.0

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1084-2014
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1084-2014. Report a data issue: hello@regdatalab.com.