Z-1088-2016 Class II Terminated

Recalled by Medtronic Navigation, Inc. — Littleton, MA

FDA device recall Z-1088-2016 was initiated by Medtronic Navigation, Inc. on Jan 29, 2016 and is designated Class II. Reason for recall: Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or… The recall status is terminated (terminated Mar 10, 2016). Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Mar 16, 2016
Initiation Date
Jan 29, 2016
Termination Date
Mar 10, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

Medtronic 0-arm 02 Imaging System Catalog Number: Bl-700-02000 The O-arm O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm O2 Imaging System is compatible with certain image guided surgery systems.

Reason for recall

Screws holding the detector panel in as a result the detector panel could make contact with other parts with in the 0-arm¿ 02 Imaging System, causing noise and vibration when being positioned or during 3D Imaging could lead to poor image quality

Distribution pattern

Worldwide Distribution to Austria only.

Code information

Serial Number: C0936

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1088-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtronic Navigation, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtronic Navigation, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtronic Navigation, Inc. →

Source: openFDA record for Z-1088-2016. Report a data issue: hello@regdatalab.com.