Z-1089-2014 Class II Terminated

Recalled by Orthofix, Inc — Lewisville, TX

FDA device recall Z-1089-2014 was initiated by Orthofix, Inc on Oct 29, 2013 and is designated Class II. Reason for recall: There is a possibility that the locking mechanism of the Azure Anterior Cervical Plate may fracture during or after implantation. The recall status is terminated (terminated Dec 9, 2014). Affected quantity: 4,172 units.

Recall details

Product Type
Devices
Report Date
Mar 5, 2014
Initiation Date
Oct 29, 2013
Termination Date
Dec 9, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,172 units

Product description

Azure Anterior Cervical Plate System, 2-Level Plate, 36 mm, Implant Grade Titanium and Nitinol, Orthofix Inc., RX Only The Azure Anterior Cervical Plate System is a temporary implant, intended for anterior fixation to the cervical spine from C2 to C7 and is indicated for: A. Degenerative disc disease, B. Spondylolisthesis, C. Trauma, D. Spinal Stenosis, E. Deformities, F. Tumor, G. Pseudoarthrosis, and H. Revision of previous surgery. The Azure Anterior Cervical Plate System is comprised of a variety of non-sterile, single use, titanium alloy and nitinol components. The system is attached to the anterior aspect of the vertebral body by means of screw to to the cervical spine.

Reason for recall

There is a possibility that the locking mechanism of the Azure Anterior Cervical Plate may fracture during or after implantation.

Distribution pattern

Worldwide Distribution: USA (nationwide) and the country of Germany.

Code information

86-01xx, 86-02xx, 86-03xx, 86-04xx, 86-05xx All lot numbers manufactured to date

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1089-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Orthofix, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1089-2014. Report a data issue: hello@regdatalab.com.