Z-1097-2014 Class II Terminated

Recalled by Ossur Americas, Inc. — Foothill Ranch, CA

Recall Details

Product Type
Devices
Report Date
March 5, 2014
Initiation Date
October 15, 2012
Termination Date
February 28, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
144 units

Product Description

The SYMBIONIC LEG is an integrated prosthetic leg for transfemoral and knee disarticulation amputees. Product Usage: SYMBIONIC LEG is intended for continuous use by amputees of low to moderate impact levels that weigh less than 125 kg for foot categories 25 30 and less than 100kg for foot categories 22, 23 and 24. The system is not intended for athletic use.

Reason for Recall

OSSUR is recalling the Symbionic Leg because it is not recognizing the low battery warning signal from the Symbionic Leg before it powers down.

Distribution Pattern

US Nationwide distribution.

Code Information

Symbionic Item Numbers: SBL12002 SBL12006 SBL12012 SBL12016 SBL12102 SBL12112 SBL12202 SBL12212 SBL12902 SBL12903 SBL12912 SBL12913 Serial #s affected: 321245, 322095, 322276, 322894; 710119-710322