Z-1101-2021 Class II Terminated
Recalled by GE Healthcare, LLC
— Waukesha, WI
FDA device recall Z-1101-2021 was initiated by GE Healthcare, LLC on Dec 10, 2020 and is designated Class II. Reason for recall: GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could… The recall status is terminated (terminated Nov 18, 2022). Affected quantity: 113 units.
Recall details
- Product Type
- Devices
- Report Date
- Feb 17, 2021
- Initiation Date
- Dec 10, 2020
- Termination Date
- Nov 18, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 113 units
Product description
GE Healthcare Precision 600FP, Classical R&F (radiographic and fluoroscopic) System - Product Usage: Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body parts.
Reason for recall
GE Healthcare has become aware of a potential safety issue associated with the improper installation of the Precision 600FP monitor suspension. A partial detachment of the monitor suspension could result in potential injury to a patient or operator.
Distribution pattern
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, WV
Code information
Serial Numbers: G1D1842032, G1D1842035, G1C17X2022, G1D1852039, Not Available*, G1D1922084, G1B16X2002, G1D1972106, G1D18X2058, G1D18Y2066, G1B1912074, G1B1772014, G1C1782016, G1D1842033, G1D18Z2071, G1D1962102, G1D1962104, G1D1862041, G1D18Y2068, G1D1972107, G1D18Z2069, Not Available*, G1D19X2118, G1D1982112, Not Available*, G1D1862043, G1D18Y2065, Not Available*, G1D1962101, G1C17X2026, G1D1832029, G1D1962105, G1B1772012, G1D1832030, G1B1772012, G1D1852038, Not Available, G1D1842031, G1C17X2023, G1C1782017, G1D1892054, G1D1932088, G1C17X2024, G1D1882049, G1D1882048, G1D1912077, G1D1932089, GID194092, G1923001923, G1793810017, G1D1882052, G1D1922080, G1D1862045, G1D1882051, G1C1792020, G1D1892055, G1D1932090, G1B1742006, G1D18Z2070, G1D1862042, G1D18X2062, G1D18X2061, G1B1762011, G1B1762010, G1D1942093, G1D1862044, G1D1882050, G1B1912079, G1D1992113, G1B1762009, G1D1852036, G1D1992117, G1D1842034, G1D1922083, G1D1952097, G1D18Z2073, G1D18X2060, G1D1912078, G1D1922085, G1D1892057, Not Available*, G1D1932091, G1D1922081, Not Available*, G1D1912075, G1D18Z2072, G1D1852040, G1D1892056, G1C1792021, G1D1982111, G1D1952100, G1B16X2005, G1D1952096, G1D1962103, G1D1922082, G1D1972108, Not Available*, G1D1932086, G1B16X2004, G1D1852037, G1D18X2059, G1D18Y2064, G1D18Y2063, G1B1942095, G1D1932087, G1D1913076, G1D18Y2067, G1D1872046, G1D1832028, G1D1992114, G1C1792018, G1C17X2025, G1B16X2003 not available* - to be supplied
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1101-2021
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about GE Healthcare, LLC →
Use it in ChatGPT
Get a free API key
How the connector works