Z-1107-2020 Class II Terminated
FDA device recall Z-1107-2020 was initiated by bioMerieux, Inc. on Dec 10, 2019 and is designated Class II. Reason for recall: Two batches of VIDAS Varicella-Zoster IgG (VZG ref 30217) may yield an invalid calibration which may delay the patient's sample result or a valid calibration may lead to a false positive patient… The recall status is terminated (terminated Feb 16, 2021). Affected quantity: 1,218 US units total (726 US, 492 OUS).
Recall details
- Product Type
- Devices
- Report Date
- Feb 19, 2020
- Initiation Date
- Dec 10, 2019
- Termination Date
- Feb 16, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,218 US units total (726 US, 492 OUS)
Product description
Reason for recall
Distribution pattern
Distributed nationwide and 9 units to Biomerieux Canada.