Z-1124-2013 Class II Terminated
FDA device recall Z-1124-2013 was initiated by Biomet U.K., Ltd. on Mar 20, 2013 and is designated Class II. Reason for recall: Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the… The recall status is terminated (terminated May 4, 2015). Affected quantity: 430.
Recall details
- Product Type
- Devices
- Report Date
- Apr 24, 2013
- Initiation Date
- Mar 20, 2013
- Termination Date
- May 4, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 430
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1124-2013
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026