Z-1148-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 4, 2015
- Initiation Date
- January 9, 2015
- Termination Date
- March 10, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 units
Product Description
NAMIC Convenience Kit, LEFT HEART KIT, UPN H7496001001111, REF/Catalog No. 600100111, STERILE, Rx ONLY For single use only. ---Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.
Reason for Recall
The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
Distribution Pattern
Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
Code Information
Lots 4476878, 4638166, 4677141.