Z-1173-2013 Class II Terminated

Recalled by DePuy Mitek, Inc., A Johnson & Johnson Co. — Raynham, MA

FDA device recall Z-1173-2013 was initiated by DePuy Mitek, Inc., A Johnson & Johnson Co. on Mar 22, 2013 and is designated Class II. Reason for recall: The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet… The recall status is terminated (terminated Feb 24, 2014). Affected quantity: 84174.

Recall details

Product Type
Devices
Report Date
May 1, 2013
Initiation Date
Mar 22, 2013
Termination Date
Feb 24, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
84174

Product description

ORTHOCORD ONE VIOLET AND ONE BLUE BRAIDED COMPOSITE SUTURE 2 (5 METRIC), Catalog Number: 223113. For use in general soft tissue approximation, and/or ligation, including orthopedic procedures.

Reason for recall

The IFU provided with the free strand ORTHOCORD sutures currently indicates that approximately 100% suture strength remains at six weeks after implantation. While this statement is true for Violet ORTHOCORD sutures, it is an inaccurate statement for Blue ORTHOCORD free strand sutures. Based on in-vivo strength retention study, a Blue ORTHOCORD sutures maintain approximately 80% of strength at six

Distribution pattern

Nationwide Distribution.

Code information

ALL from Jan 1 2010 - Corrected IFU implementation date

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1173-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1173-2013. Report a data issue: hello@regdatalab.com.