Z-1175-2015 Class II Terminated

Recalled by Navilyst Medical, Inc — Glens Falls, NY

FDA device recall Z-1175-2015 was initiated by Navilyst Medical, Inc on Jan 9, 2015 and is designated Class II. Reason for recall: The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit… The recall status is terminated (terminated Mar 10, 2017). Affected quantity: 5 units.

Recall details

Product Type
Devices
Report Date
Mar 4, 2015
Initiation Date
Jan 9, 2015
Termination Date
Mar 10, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 units

Product description

NAMIC Convenience Kit, LEFT HEART KIT, UPN H749601506721, REF/Catalog No. 60150672, STERILE, Rx ONLY ---For single use only. --- Device Listing Number D136588. Product Usage: NAMIC Convenience Kits are intended to be used in Fluid Management and/or Invasive Pressure Monitoring Systems.

Reason for recall

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.

Code information

Lots 4663410 and 4664414.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1175-2015
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Navilyst Medical, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Navilyst Medical, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Navilyst Medical, Inc →

Source: openFDA record for Z-1175-2015. Report a data issue: hello@regdatalab.com.