Z-1176-2014 Class II Terminated

Recalled by Biosense Webster, Inc. — Irwindale, CA

FDA device recall Z-1176-2014 was initiated by Biosense Webster, Inc. on Feb 20, 2014 and is designated Class II. Reason for recall: Biosense Webster is recalling the PentaRay Nav Catheter because it has the potential to have partial tip separation from the catheter shaft. Biosense Webster is expanding the recall on 5/1/14 to… The recall status is terminated (terminated Oct 24, 2014). Affected quantity: 15,672 units total (12,804 units in US).

Recall details

Product Type
Devices
Report Date
Mar 19, 2014
Initiation Date
Feb 20, 2014
Termination Date
Oct 24, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15,672 units total (12,804 units in US)

Product description

PENTARAY NAV High-Density Mapping Catheter and PENTARAY NAV ECO Catheter, Catalog No. D128201, D128202, D128204, D128205, D128207, D128208, D128210, D128211. The Biosense Webster PENTARAY NAV High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart.

Reason for recall

Biosense Webster is recalling the PentaRay Nav Catheter because it has the potential to have partial tip separation from the catheter shaft. Biosense Webster is expanding the recall on 5/1/14 to include all lots of the PentaRay Nav Catheter and the PentaRay Nav ECO Catheter because they can lead to an insufficient bond of the distal tip to the catheter shaft.

Distribution pattern

US Nationwide Distribution.

Code information

Lot Numbers: All Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1176-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Biosense Webster, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1176-2014. Report a data issue: hello@regdatalab.com.