Z-1199-2019 Class II Ongoing
FDA device recall Z-1199-2019 was initiated by GE Healthcare, LLC on Mar 15, 2019 and is designated Class II. Reason for recall: There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal… The recall status is ongoing. Affected quantity: 1 device.
Recall details
- Product Type
- Devices
- Report Date
- Apr 24, 2019
- Initiation Date
- Mar 15, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1 device
Product description
Reason for recall
Distribution pattern
Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.