Z-1200-2019 Class II Ongoing
Recalled by GE Healthcare, LLC
— Waukesha, WI
FDA device recall Z-1200-2019 was initiated by GE Healthcare, LLC on Mar 15, 2019 and is designated Class II. Reason for recall: There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal… The recall status is ongoing. Affected quantity: 49 devices.
Recall details
- Product Type
- Devices
- Report Date
- Apr 24, 2019
- Initiation Date
- Mar 15, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 49 devices
Product description
Innova IGS 630, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
Reason for recall
There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Distribution pattern
Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.
Code information
Model Number 5730860; System ID (Serial Number, UDI Number): 352333VASC3 (B3-17-006, 01008406821086381117020021B3-17-006); 727869SPECIALS (B3-18-005, 01008406821086381118020021B3-18-005); 407518BIPLANE (B3-17-005, 01008406821086381117010021B3-17-005); 305227KBP630 (B3-16-017, 01008406821086381116080021B3-16-017); 401BIPLANE1 (B3-18-001, 01008406821086381118010021B3-18-001); 941629IGS630 (B3-18-002, 01008406821086381118010021B3-18-002); 505841CATH3 (B3-17-032, 01008406821086381117120021B3-17-032); 7022553131 (B3-17-011, 01008406821086381117030021B3-17-011); 7028803131 (B3-17-023, 01008406821086381117080021B3-17-023); 615769SP3 (B3-17-013, 01008406821086381117040021B3-17-013); 214618INIGS (B3-16-029, 01008406821086381116120021B3-16-029); 281890BP (B3-17-024, 01008406821086381117090021B3-17-024); 956632IGS630 (B3-17-029, 01008406821086381117120021B3-17-029); 804320IGS630 (B3-16-016, 01008406821086381116070021B3-16-016); DZ1009VA03 (B3-17-015, 01008406821086381117070021B3-17-015); 220826RX10 (B3-17-003, 01008406821086381117010021B3-17-003); 082416040099 (B3-19-001, 01008406821086381119020021B3-19-001); 082416160088 (B3-18-023, 01008406821086381118110021B3-18-023); 082416100170 (B3-19-003, 01008406821086381119020021B3-19-003); XV8624CO (B3-18-012, 01008406821086381118050021B3-18-012); N4192956 (B3-17-009, 01008406821086381117030021B3-17-009); A5820150 (B3-16-023, 01008406821086381116100021B3-16-023); YV1920 (B3-16-020, 01008406821086381116100021B3-16-020); YV1922 (B3-17-002, 01008406821086381117010021B3-17-002); YV1921 (B3-16-027, 01008406821086381116120021B3-16-027); YV4500 (B3-16-032, 01008406821086381117010021B3-16-032); YV4501 (B3-17-027, 01008406821086381117110021B3-17-027); YV2503 (B3-16-019, 01008406821086381116090021B3-16-019); YV4503 (B3-18-003, 01008406821086381118020021B3-18-003); YV4502 (B3-17-030, 01008406821086381117120021B3-17-030); 0850060758 (B3-17-014, 01008406821086381117070021B3-17-014); 0850060755 (B3-17-001, 01008406821086381117010021B3-17-001); 0850060759 (B3-18-013, 01008406821086381118060021B3-18-013); KW1008VA01 (B3-17-019, 01008406821086381117070021B3-17-019); KW1005VA02 (B3-16-025, 01008406821086381116110021B3-16-025); KW1005VA01 (B3-16-026, 01008406821086381116110021B3-16-026); LY1067VA01 (B3-17-010, 01008406821086381117030021B3-17-010); PK1003VA01 (B3-17-018, 01008406821086381117060021B3-17-018); RU2011VA02 (B3-17-004, 01008406821086381117010021B3-17-004); RU9774VA01 (B3-16-021, 01008406821086381116100021B3-16-021); SA2513VA01 (B3-16-024, 01008406821086381116100021B3-16-024); ZA2940VA01 (B3-17-025, 01008406821086381117100021B3-17-025); SD1084VA01 (B3-18-010, 01008406821086381118050021B3-18-010); 06189VAS06 (B3-17-028, 01008406821086381117110021B3-17-028); 55001VAS02 (B3-16-031, 01008406821086381116120021B3-16-031); 5973XR0085 (B3-17-007, 01008406821086381117020021B3-17-007); 5973XR0075 (B3-16-030, 01008406821086381116120021B3-16-030); 5973XR0076 (B3-17-008, 01008406821086381117030021B3-17-008); 5973XR0094 (B3-17-012, 01008406821086381117040021B3-17-012)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1200-2019
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about GE Healthcare, LLC →
Use it in ChatGPT
Get a free API key
How the connector works