Z-1201-2019 Class II Ongoing

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-1201-2019 was initiated by GE Healthcare, LLC on March 15, 2019 and is designated Class II. Reason for recall: There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneo… The recall status is ongoing. Affected quantity: 22 devices.

Recall Details

Product Type
Devices
Report Date
April 24, 2019
Initiation Date
March 15, 2019
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
22 devices

Product Description

Innova IGS 6, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.

Reason for Recall

There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.

Distribution Pattern

Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.

Code Information

Model Number 5771387; System ID (Serial Number, UDI Number): 501257IGS630 (B3-18-014, 01008406821246141118070021B3-18-014); 951486RU630 (B3-18-009, 01008406821246141118040021B3-18-009); 714993IGS630 (B3-18-004, 01008406821246141118020021B3-18-004); 906225EP1 (B3-18-015, 01008406821246141118070021B3-18-015); 906225CATH4 (B3-18-016, 01008406821246141118080021B3-18-016); 843797IGS30 (B3-18-030, 01008406821246141119010021B3-18-030); 4603984XV41 (B3-18-019, 01008406821246141118090021B3-18-019); MCARIGS630 (B3-18-008, 01008406821246141118040021B3-18-008); 509473RM1B (B3-18-028, 01008406821246141119010021B3-18-028); 414649IGS8 (B3-18-020, 01008406821246141118090021B3-18-020); HU1099VA07 (B3-18-024, 01008406821246141118110021B3-18-024); A670060221 (B3-19-002, 01008406821246141119020021B3-19-002); A5163326 (B3-18-007, 01008406821246141118030021B3-18-007); YV4504 (B3-18-018, 01008406821246141118090021B3-18-018); YV4505 (B3-18-027, 01008406821246141118120021B3-18-027); YV4507 (B3-18-022, 01008406821246141118110021B3-18-022); 0850060762 (B3-18-025, 01008406821246141118110021B3-18-025); 0850060763 (B3-18-026, 01008406821246141118120021B3-18-026); KW1155VA02 (B3-18-017, 01008406821246141118080021B3-18-017); PL1689VA01 (B3-18-021, 01008406821246141118100021B3-18-021); PL0302VA02 (B3-18-011, 01008406821246141118060021B3-18-011); 06189VAS11 (B3-18-029, 01008406821246141119010021B3-18-029)