Z-1203-2017 Class II Terminated

Recalled by Merge Healthcare, Inc. — Hartland, WI

FDA device recall Z-1203-2017 was initiated by Merge Healthcare, Inc. on May 31, 2015 and is designated Class II. Reason for recall: Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure. The recall status is terminated (terminated Mar 13, 2020). Affected quantity: 341 sites potentially have the affected product versions.

Recall details

Product Type
Devices
Report Date
Feb 22, 2017
Initiation Date
May 31, 2015
Termination Date
Mar 13, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
341 sites potentially have the affected product versions

Product description

The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the Merge Hemo software.

Reason for recall

Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure.

Distribution pattern

Distribution was nationwide to medical facilities. There was also military and government distribution.

Code information

Patient Data Module (PDM) V2 model: Serial numbers 10.03.XXXXX

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1203-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1203-2017. Report a data issue: hello@regdatalab.com.