Z-1203-2017 Class II Terminated
FDA device recall Z-1203-2017 was initiated by Merge Healthcare, Inc. on May 31, 2015 and is designated Class II. Reason for recall: Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure. The recall status is terminated (terminated Mar 13, 2020). Affected quantity: 341 sites potentially have the affected product versions.
Recall details
- Product Type
- Devices
- Report Date
- Feb 22, 2017
- Initiation Date
- May 31, 2015
- Termination Date
- Mar 13, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 341 sites potentially have the affected product versions
Product description
The Schiller PB 1000, also known as: Merge Hemo Patient Data Module (PDM), V2 model, a major component of the Merge Hemo software.
Reason for recall
Non-invasive blood pressure pumps (NIBP) voltage is not sufficient and causes the NIBP pumps to fail prior to reaching adequate initial pressure.
Distribution pattern
Distribution was nationwide to medical facilities. There was also military and government distribution.
Code information
Patient Data Module (PDM) V2 model: Serial numbers 10.03.XXXXX