Z-1225-2017 Class II Terminated

Recalled by Tosoh Bioscience, Inc. — South San Francisco, CA

FDA device recall Z-1225-2017 was initiated by Tosoh Bioscience, Inc. on Nov 7, 2016 and is designated Class II. Reason for recall: The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of… The recall status is terminated (terminated Feb 9, 2021). Affected quantity: 682 analyzers.

Recall details

Product Type
Devices
Report Date
Mar 1, 2017
Initiation Date
Nov 7, 2016
Termination Date
Feb 9, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
682 analyzers

Product description

Tosoh Bioscience HLC-723G8 Analyzer for HbA1c testing.

Reason for recall

The device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the diagnostic medical decision point of 6.5%.

Distribution pattern

Worldwide distribution. The products were shipped to the following states: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, AND WY. The recalled products were shipped to the following countries: Brazil, Chile, Columbia, Guatemala, Costa Rica, Dominican Republic, Eduador, Nicaragua, Panama, Peru, Puerto Rico, Uruguay,

Code information

All active analyzers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1225-2017
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tosoh Bioscience, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1225-2017. Report a data issue: hello@regdatalab.com.