Z-1227-2022 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 15, 2022
- Initiation Date
- April 18, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 88,379 batteries
Product Description
Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstr¿om Carestation and Engstr¿om PRO Ventilators.
Reason for Recall
Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.
Distribution Pattern
Worldwide Distribution: US (nationwide) to AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI and OUS (internationally) to Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Cote D'Ivoire, Mexico, Uruguay, United Kingdom, United Arab Emirates, Turkey, Thailand, Tanzania, Sweden, Spain, Singapore, Saudi Arabia, Russia, Romania, Portugal, Poland, Philippines, Panama, Omar, Norway, Nigeria, Nicaragua, New Zealand, Netherlands, Malaysia, Latvia, Kuwait, Korea, Kenya, Kazakhstan, Japan, Italy, Ireland, Indonesia, India, Hungary, Germany, France, Finland, Estonia, Egypt, Ecuador, Denmark, Czech Republic.
Code Information
Field Replacement Unit kits PN: 1009-5682-000-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V) and PN: 5856787-S (BTRY SEALED LEAD ACID RECHARGEABLE 12V PAIR).